|
SAFETY
IN THE ELECTRON
MICROSCOPE
ROOM
by
S. K. Chapman
Keywords: Electron
microscopes, safety, maintenance
MICROSCOPY AND ANALYSIS, MARCH 1989
Steve
Chapman has been involved with electron microscopes since 1964.
Initially working as a service engineer in the UK his service activities
eventually took him throughout Europe.
Moving into sales and marketing he worked with a number of
electron microscope manufacturers where he was probably most at home in
a teaching or application environment. In 1981 he set up his own training organisation Protrain,
adding a service company to his repertoire in 1983.
Running courses in many parts of the world, north and south of
the equator, he has written a number of books on the operation and
maintenance of scanning and transmission electron microscopes. However,
Protrain is not only involved in electron microscopy but also in motor
sport, where their (Go) kart racing courses are a considerable success.
Many
texts have been written on the use of the scanning and transmission
electron microscopes. Our own Royal Microscopical Society has through
its safety committee produced a considerable amount of useful literature
on the hazards that are faced in the preparation of the specimen.
However, for every instrument there is invariably a person who is
involved with day to day maintenance of that instrument.
Very few papers are written on their safety and as a guide to the
safe procedures for maintaining the instrument.
In this article an outline is given of the problems that may
confront the microscopist carrying out even the most basic maintenance
procedures. The hints and
tips are designed to make working with the instruments a much safer
occupation.
Whilst
the actual electron microscope room is probably one of the safest
areas in an electron microscope laboratory, it contains a number of
hazards that demand attention. Some
of these are present on all occasions.
Whilst others only present themselves when some form of
maintenance or other routine procedure is being performed.
We
are all aware that on opening an electron gun that has been in use, the
cathode assembly will be very hot.
However, there are other hazards that should be considered prior
to this point. Perhaps at this stage we should set a standard for any
action toward, or performed around, the microscope.
"Never attempt any routine maintenance task without thinking
of the possible hazards involved with that task!"
Most
manufacturers include a "shorting bar" which falls across the
cathode assembly as the gun is opened.
This bar under normal circumstances should never pass a current.
Only if things go wrong could it save your life.
In a modern microscope (post 1970) two actions need to be omitted
from the conventional procedures for the high voltage to remain applied
to the cathode when the system is at air.
A vacuum trip should immobilise the high voltage supply when air
is let into the gun. If you
have also forgotten to switch off the high voltage, a combination of
these two mistakes could lead to a dangerous situation.
We can see that a modern instrument should protect us in these
two ways, however, failing to follow conventional switching procedures
would leave an operator open to considerable danger.
It is true that at a high accelerating voltage most people would
smell the ozone from the gun arcing before completely opening the
electron gun chamber, but if you don't want to let a problem get that
far the message is "do not take short cuts."
If
you use an elderly microscope (pre 1970) that does not use a high
frequency generator, the high voltage is generated with the assistance
of a very large capacitor. This
is usually shorted to earth within the tank when the high voltage is
switched off. I have had
the frightening experience of shorting out this type of gun with a
screwdriver (Figure 1). Only the blinding flash indicating that the
internal shorting bar had failed to discharge the capacitor; it had
fallen off! Always be sure
that the gun is shorted out. Always
work as if safety devices have failed and never cut corners.
Now
for that hot cathode. Wear
gloves that are thick enough to protect you from the hot metal. Chamois leather gloves are ideal and they will not
contaminate the gun chamber. The
filament has been up at about 25000C so no wonder the
assembly is very hot! Another
problem in this area is reaching the gun itself.
Many instruments provide steps for the operator to have access to
the gun level. If you have
not had steps provided with the instrument, do not perform a balancing
act on a chair or stool. Please
purchase a pair of steps as you will be changing a filament hundreds of
times through the life of an instrument.
Another
area of gun maintenance may require the chamber to be removed for
cleaning. Please be advised
by your manufacturer on how this should be accomplished, and about the
weight of the chamber. Take
great care because one mistake may mean a damaged chamber or even a
damaged operator.
The
electron gun is also a source of x-rays.
If the gun or condenser area is dismantled for any reason, you
must ensure that it is checked for radiation leaks prior to using the
instrument. Whilst all
engineers will rebuild the column with great care, only the smallest of
misalignments could result in a radiation leak.
This precaution is not necessary after the opening of a gun for a
filament exchange, it is only applicable after the column units have
been taken apart. Whilst on
this subject, do not run the microscope with any of the column shroud
panels removed. The
manufacturer may be using these to provide extra radiation protection.
The
specimen chamber of the scanning electron microscope is also an area
where you may inadvertently come into contact with a high voltage.
Once again this problem will only occur if a number of safety
devices have failed. Having
said that, it is "quite exciting" to see a one centimetre long
spark arcing across from the detector to the bottom of the final lens!
There will be up to ten thousand volts on the secondary electron
detector when it is in operation. When
admitting air to the chamber the detector high voltage should be
automatically switched off, if not - sparks!
There is nothing that you can do except turn off the detector if
the instrument has the appropriate switch.
So take care and do not place your hand near to the detector
under normal circumstances. If
there is a problem you will hear the detector arcing, and if you do,
close the door and pump the instrument down.
A call for service is your next task!
Many
instruments use liquid nitrogen as a coolant for an anticontamination
device or an energy dispersive x-ray detector.
When handling any very cold liquid great care must be taken.
Small droplets of liquid nitrogen falling on the skin will
rapidly evaporate. The
greatest danger is to the eyes, they should always be protected whilst
handling the very cold liquid. Always
try to stand well above the container into which the liquid is being
poured.
The
EDS detector also becomes a hazard in itself.
The position of the detector, particularly on a transmission
electron microscope, makes it a prime cause of cracked heads.
Try to protect your operators by padding the corners of the
detector support. Filling the detector on a scanning electron microscope
may require moving behind the instrument.
Take care not to tread on any cables or the pumping lines.
The
rotary pump seems to be an innocuous part of the electron microscope
laboratory but it contains hidden dangers!
The fumes from any hot oil are said to be carcinogenic (Ref 2),
contaminated vapours even more so.
Therefore it is most important that the rotary pump is either
fitted with a good quality oil mist filter, or its fumes are ducted to
the outside world. Keep a
constant check on the state of a filter.
If it becomes discoloured or moist with oil, change it!
If you can smell oil when pumping out a vessel, check the filter
and the state of the oil. Do
not forget that ALL the rotary pumps in your laboratory should be
treated in this way. In particular scanning electron microscopists should pay
attention to the pump on their sputter coater. This pump is constantly
pumping against the gases given off by specimens and their adhesives. It will certainly be the most contaminated of your rotary
pumps. Even a good filter
may be inefficient when the pump oil is severely contaminated, so be
sure to have the pump oils changed at least every year.
More often if they are heavily used.
Fresh oils are either colourless or a faint yellow colour; when
contaminated they become dark yellow to brown.
Do not get the contaminated oil on your hands if you carry out
your own pump maintenance, as it may also cause dermatitis. If this dirty task is left with your engineer make sure that
it IS being carried out. If
you are not being charged for the oil it is probable that it is not
being changed! It is your
safety that is at stake, find out.
A
most recent problem for me is the sensitising of this electron
microscope operator by cleaning agents.
Even that most user-friendly material, like Duraglit, will start
to affect you if you use it year after year.
Twenty years of cleaning electron microscopes and now the skin
peels off my finger tips when I have been using my favourite cleaning
media. What will affect
you? Treat all the cleaning
agents as a hazard and you will not be far wrong.
Watch out for the vapours from acetone, and ether, or any other
solvent that you may use. This
is a particular problem if you are using a solvent as a leak-detecting
agent. In this case the
solvent is sprayed at possible leak sites, an indication of it been
drawn into the vacuum system being a dramatic rise in pressure on a
Penning gauge. This is an
area that you should be aware of when the service engineer is looking
for leaks also. It does not take long for the air in most electron microscope
rooms to become saturated with the vapour.
Keep the doors open and visit the engineer from time to time.
The
ultrasonic cleaner is a great asset in any laboratory but it too is a
hazard. We all hate the
constant high pitched noise emitted from the unit, but the real problem
comes with holding or handling of components in the cleaner.
Do not take components from, or hold components in, the cleaner
whilst it is switched on. Damage may be caused to the wrists.
Do not ask me what goes wrong but it makes the wrists very
painful if you ignore this advice.
Keep all fluids covered whilst you are using them for cleaning.
The heat generated by the cleaning action will cause a high level
of vapourisation. Keep the
ultrasonic cleaner in a fume cupboard if at all possible.
The
annual service visit, or that panic emergency call, does not mean that
you have a chance to dive off to the library for a day or so.
Stay with your service engineer.
There are many dangers involved with the service of an electron
microscope. Anyone removing
panels from an instrument has the possibility of coming into contact
with an electrical circuit. Do
not leave a service engineer with an instrument without someone within
shouting distance. Safety
in YOUR laboratory is YOUR responsibility!
Should
you fall into the category of being totally your own service engineer
may I offer a few additional words of warning.
You may conduct your mechanical maintenance with a high degree of
skill, thinking well ahead of what the next action may be.
Good,
then you should be safe. However,
how often do you venture into the electrics and electronics?
You may well feel capable of changing fuses, do you make sure
that the instrument is switched off?
Do you remove any metal objects from your hands and wrists? They could be the cause of you getting an electric shock!
Do you make sure that the rotary pump is switched off before
commencing maintenance (belt burns hurt) and do you take great care when
working round a hot diffusion pump?
Watch out when handling the water connections. Switching off the microscope first is a good starting point.
One of the most dangerous areas with many microscopes is the room
light switching circuit. Make sure you know exactly which area is switching off when
you switch off the microscope. Some
installations allow the mains for the room light to remain connected
even when the microscope is switched off; does yours?
If you are receiving guidance over the phone double-check the
action recommended. If you
are guided into the electrics and electronics make sure that you know
EXACTLY what action you are to take.
When using a multimeter make sure that the settings are EXACTLY
those the engineer has suggested; AC, DC, volts or ohms. Make sure the connections that you make are EXACTLY those
suggested. If in doubt ASK
AGAIN. Finally, is there
always someone within shouting distance when YOU are acting as service
engineer? Well there should
be! Please remember that
advice over the phone is invariably free.
Do not be afraid to ask and if you have any doubt about tackling
a problem, DON'T.
Safety
in the electron microscope room, as I have already said, is probably
considered to be the lesser of your laboratories evils.
Do not fall into the trap of becoming careless, because that is
when the electron microscope is at its most dangerous!
References:
Chapman,
S.K., Maintaining & Monitoring The Transmission Electron Microscope,
Oxford University
Press,
1986 pp 37-39.
Meek,
G.A., Practical Electron Microscopy For Biologists, John Wiley and Sons,
1977, pp 4546
|