|
A.
The size of the probe in the SEM has an affect in a number of
areas –
1)
The
ultimate imaging resolution.
2)
The
level of imaging signal from a sample.
3)
The
level of x-ray signal from the sample.
4)
The
relative levels of secondary and backscattered electrons emitted from
the specimen; larger spots result in more backscattered electron
influence in the image.
B.
The actual probe size on the surface of the specimen relates to a
number of other settings within the instrument
1)
The
position of the filament in the cathode (shorter distances give a higher
emission level allowing a higher bias to be applied reducing the source
size whilst retaining a usable emission level).
2)
The
applied bias level (as above).
3)
The
accelerating voltage (the extraction of electrons from the cathode is
less efficient as the kV is lowered due to the anode to cathode distance
being retained at that desired for the highest instrument kV.
The distance should be 1mm for every 2 kV for maximum gun
efficiency, therefore in theory the anode should be raised as the kV is
lowered.)
4)
The
setting of the condenser lenses (A high strength first condenser
provides a smaller spot.)
5)
The
focal length (WD) of the final lens. (A longer focal length requires a
weaker lens to focus and as a result a higher level of spherical and
chromatic aberration. A
longer focal length also results in more outside interference from
electrical fields in the laboratory.)
C.
The ultimate resolution of an instrument depends upon the
settings mentioned in “B” above when a test specimen is the subject.
This specimen should consist of an artificial structure
containing heavy particles, usually gold, on a light element support,
usually carbon or latex, this configuration of “something on
nothing” exaggerates performance but may provide a resolution standard
for most instruments. In
reality the resolution attained with a day to day specimen depends upon
that specimen’s ability to emit electrons, for example carbon based
materials provide a weaker signal.
D.
The size of the beam spot, should it require measurement, may be
determined using the edge affect method.
1)
A
disk aperture is placed in the microscope sitting over a hole in a stub.
2)
The
magnification is increased to maximum and the edge of the aperture
brought to focus at the desired probe size.
3)
A
line scan is run across the edge of the aperture.
4)
A
wave form is produced from the line scan where a measurement is taken
from 10%* horizontally into the slope up to 90%* horizontally into the
slope, this figure is related to the magnification and is said to be the
actual probe size. (check these figures *)
E.
The performance of the microscope is better checked through the
use of a sample as described in “C”.
Periodic tests where the image quality is related to the previous
test are sufficient for most laboratories, but point to point
measurements on the structures may also be made.
If measurements are taken a magnification calibration will also
be required. Great care
should be taken to ensure that the magnification calibration is
performed under identical conditions, including spot size, to that of
the resolution test. Changing
the spot size on many older instruments changes the magnification due to
a change in final lens focal length.
|