Contamination Rate - contamination
is a cracking of vapours within the vacuum by the heat of the beam on a
surface, hydrocarbons and silicons being deposited hence the low signal
level dark lines or rectangles. SEM
contamination rate is very much specimen dependant but by taking a
constant approach this may be a useful test.
I use sputter coated latex spheres the specimen being in the
microscope one hour prior to the test.
A typical rough tough microscope used without any care gives
10nm/min over my 20 minute test period.
Under similar (emphasise similar) conditions a well kept air locked
instrument will come down to 2.5nm/min.
Add a cold finger around the final lens similar to that used in a
cryo system and you are down to <1.5nm/min.
Drift Rate - I thought SEM stages were very poor however testing a
good number of instruments over a wide price range I found that over a
twenty minute period the amount of drift was less than the instruments
resolution, in other words the sample did not move.
If it did I always found an earth problem not a stage drift
problem. I no longer bother with this test unless I have a worry about a
particular stage stability.
Most
of my work has been on run of the mill instruments with the best results
from the modern twin detector FEG systems. In these instruments a good
cold finger sitting around the final lens is the difference between good
and amazing results - contamination IS the killer of high resolution
microscopy in my mind.
Vacuum
System.
You should know the pump down time from air, and the time each pump
is operating during the pump down procedure.
These values should be obtained for each air locked area of the
instrument. The exact valve
changes during each pump down cycle should be documented if a mimic
display is provided.
ROUTINE
MAINTENANCE PROCEDURES
During
use the tungsten filament will thin, and eventually break, due to
evaporation and oxidation. Filaments will break in the following fashion:-
1.
A normal break occurs to one side of the tip, the break should be
between two tapered ends.
2.
An overheated filament will break as above, except the ends will
not taper, they will either be in the form of blobs of metal, or if the
blobs have fallen off, two blunt ends.
3.
A filament effected by a very poor vacuum will break due to severe
oxidation. The filament will
seem to break with the normal taper, but its life will have been very
short. You will notice how
the filament is very thin, far thinner than for a normal break.
Even if the gauge indicates the vacuum is good, this is not often a
true indication of the gun vacuum, as the gauge is normally positioned
near to the high vacuum pump.
The
filament ceramic also acts as
an indication of the filament's life.
1.
If the filament has been carefully saturated, not overheated, the
base will be a light blue in colour.
The colour comes from a light coating of evaporated tungsten.
2.
If the filament has been run very hard, either due to overheating
in error, or to the filament being placed very close to the cathode, in
order to obtain improved emission levels, the ceramic will be a dark blue
in colour.
3.
An orange to brown colouration is due to contamination, the
filament had been operating in a poor vacuum environment. This colouration
would be expected if the filament had failed due to excess oxidation.
The
cathode assembly is made of
stainless steel, which when dirty may be coated with tungsten (blue to
black in colour), and general contamination from the vacuum (orange yellow
colour). Prior to fitting a
new filament the components must be polished perfectly clean.
A number of cleaning methods are available, the choice being
whether to clean by hand or to use an ultrasonic cleaner.
Cleaning
by hand,
you may use any metal polishing media that is suitable for brass or
stainless steel. The most
important point is, to remove ALL of the polishing media, when the
cleaning process is complete. Each
cleaned component should be stored in a lint free tissue or aluminium foil
until ready for use. Prior to
fitting a new filament, each component should be checked with a hand lens,
to ensure there are no debris remaining.
The
o-rings that are disturbed,
when entering the gun chamber, should be washed in hot soapy water.
The o-ring seat is best cleaned with a solvent prior to refitting
the ring. Do NOT grease the
rings. The only o-rings that
require grease are those that make a moving seal, e.g. aperture drives,
and mechanical gun alignment seals. Grease
is the largest contributor to vacuum contamination, the less grease you
use the better the column and specimen contamination rates will be.
The
electron column will require
cleaning, in part, or as a whole, under the following circumstances.
1.
If you are unable to correct the objective astigmatism, in the
first case the final aperture should be changed.
2.
If the above action does not remove the problem, then the aperture
holder should be cleaned.
3.
Should the problem remain after following the above two
sections, then the remainder of the column liner should be cleaned, and
the apertures cleaned or replaced
When
cleaning the aperture holder or column liner tubes, any of the techniques
suggested for cleaning the cathode assembly should be more than adequate.
In the SEM case it does no harm placing a number of components in
an ultrasonic cleaner, the external surfaces are of no consequence.
Thin
metal apertures,
usually silver in colour, require cleaning techniques that use heat to
drive off the contamination. Two
metals are used for thin metal apertures, platinum or molybdenum.
If the manufacturer's instruction book does not have details of
which type of aperture are fitted to your microscope, there is a simple
test. Platinum apertures may
be heated in a Bunsen flame, but molybdenum apertures will turn black.
To
clean a platinum aperture you
will need to purchase a small platinum boat, or a pair of platinum tipped
tweezers. The aperture is
placed on the boat, or held in the tweezers, and heated in a flame until
it is orange to red heat. Check
the aperture under a microscope and if it does not look perfectly clean
repeat the procedure. Molybdenum
apertures require heating in a high vacuum environment, to prevent
oxidation. A high vacuum evaporating system, with a molybdenum boat linked
between two low voltage terminals, is ideal.
The aperture is heated to orange-red heat until the dark patch, the
contamination, is removed. After
cleaning apertures of any material, they should be checked with a
microscope, to see that they are both perfectly clean, and perfectly
round. Throw away apertures
that do not respond to cleaning, or become distorted by the cleaning
procedure.
The
SEM specimen area should be
cleaned each week with a vacuum cleaner, and the door o-ring checked for
debris or damage. If the
specimen stubs fit into a removable cup, or cups, these should be cleaned
with a solvent. For
both SEM and TEM do try to encourage operators to wear light weight gloves
when handling specimens. They
cut down the transfer of finger grease and considerably reduce both
specimen and column contamination.
The
vacuum system of the SEM should not be ignored when routine
maintenance is being considered. The first priority with any rotary pump is to remove the oil
fumes. The cheapest technique
is to vent the pump outlet to the outside world through the use of a
plastic pipe. If your
microscope is deep within a building you should use a filter fitted to the
pump exhaust. The fluid in
the rotary pump works very hard and therefore contaminates rather rapidly.
If you have a pump filled with one of the very expensive low
contamination oils, drain and refill the pump every two years.
If you use a pump that contains a conventional fluid, change this
every year. A pump that is
constantly being used to pump a system from air takes in a high level of
moisture. The initial surge of other vapours from the specimen.
These contaminants gradually reduce the efficiency of the pump
fluid, the pumping speed is reduced, and the exhaust emits rather nasty
vapours. Three very good
reasons for regularly changing the fluid.
Diffusion
pumps
require no maintenance if correctly operated.
The pump is most damaged by condensation, moisture building up on
its outer surface. This leads
to corrosion, water leaks, and damaged heaters.
Diffusion pump should be operated with a water temperature within
two degrees of room temperature. The best evaluation of the pump
temperature is to check the temperature of the SECOND COIL from the base
of the pumps. This should be
very slightly warm to the touch. Vary
the water flow until the correct temperature is attained. The only reason
for changing the fluid in the diffusion pump is if the pumping speed is
being impaired. Check that
your microscope does fail-safe by switching off the water once a month.
Fail safe devices may jam and cause an expensive pump melt down!
The
general vacuum performance of the microscope, and therefore the
contamination rate, may be improved through the use of a nitrogen bleed
system. If a dry nitrogen gas
cylinder is connected with the air inlet for the column, with a very low
pressure setting, the dry gas will preserve a much cleaner column
environment. Under these
circumstances, filament life, column cleanliness, pump down time, and
specimen contamination rate all improve.