Contrast in SEM Images

 

Whilst it is understandable that the surface topography of a specimen dictates the contrast changes that portray the specimen image, there are other reactions that add or subtract contrast from the image.  Contrast changes will result from a change in accelerating voltage, spot size (probe current) and tilt angle, each of which change the secondary electron to backscattered electron ratio.  It should be noted that the emission of secondary electrons from a given probe current will remain relatively constant as the accelerating voltage is changed.  However as the accelerating voltage is increased the BSE influence increases, diminishing the contribution to the image from the SE.  Another area of influence is the design of the microscope, the detector position, the lens shape and the chamber shape may all have a considerable influence upon the image contrast.  

The image produced by the Everhart-Thornley detector, is made up of a number of electron signals.  Not only do secondary electrons become drawn into the detector, but also the high-energy backscattered electrons entering the detector, react with the scintillator in exactly the same way as the secondary electrons.  The detector is unable to discriminate between electrons of differing energy, as they strike the scintillator, in either case photons are produced and processed by the photomultiplier.  In truth although secondary electrons dominate the signal entering the detector the majority of these electrons have been generated though converted backscatter, thus the dominating reaction in most images is that which produces backscattered electrons.  With many instruments it is very difficult to eliminate these signals only on the most sophisticated instruments may “pure” secondary electron images be produced.

Whilst as we have mentioned above the general belief is that SE are the originators of the image quality when it is actually the influence of BSE that adds higher contrast, greater image depth and shadows to the image, often making the image more “exciting” and informative.

1     Signal collection

Scanning electron microscopes have developed in many ways but by far the most innovative progress has been made in the way that the electron signal is detected.  Whilst retaining the basic Everhart-Thornley (E-T) detector two styles of signal collection have been developed.

 1.1     Collection from within the specimen chamber

 The conventional position for the E-T detector is within the specimen chamber.  In this case the signal variation, or contrast, may be changed not only as described earlier, but also through the vertical movement of the specimen within the specimen chamber.  Moving the specimen in this manner varies the signals reaching the E-T detector adding or subtracting the BSE influence.  Not all of this change is due to line of sight backscatter, a great deal is due to converted backscatter from components within the specimen chamber.  The high energy backscatter spray away from the specimen surface in all directions.  The BSE striking components of the chamber with energies near to the accelerating voltage react with these surfaces in the same way as the electron beam with the specimen.  Secondary electrons, backscattered electrons and x-ray are all produced from the interaction of the BSE with the chamber components.  Depending upon the ease that these electron signals “see” the E-T detector they will add their contribution.  SE will be attracted into the E-T detector and BSE may reach the detector through direct line of sight or through multiple scattering reactions.  It is important to remember that SE produced by BSE striking a surface actually carry the information that the BSE brought from the specimen.

 The manufacturers detector design may also have an influence on which electrons reach the detector.  Small scintillators reduce the BSE content simply by offering a smaller surface area with which to react.  However as the secondary electrons are actually attracted into the detector their number is not related to scintillator size.  Some manufactures have an open mesh collector offering no signal discrimination, whilst others have a cone or shield on the front or around the detector that constrains the detected signal.  The position of the detector will have an influence upon the signals it collects and under which conditions high levels of SE or BSE collection take place.

  1.2  Double detector systems

 Since the early 1980s instruments with two E-T detectors have been commercially available.  The detectors are mounted one within the specimen chamber, the other being placed above the final lens.  In this configuration the final lens acts as an electron filter, the lens field and lens geometry preventing the upper detector having line of sight BSE influence.

 High resolution imaging of areas within large specimens had been prevented in conventional instruments as moving the large sample nearer to the final lens prevented most of the signal from reaching the E-T detector; the specimen was in the way!  With a detector above the final lens such problems do not exist and the following reactions influence signal collection.

The electron beam strikes the specimen with the normal beam-specimen reactions.  The low energy SE are unable to break away from the lens axis due to the lens field being sufficient to control the beam and much too strong to allow the SE to escape.  These electrons are therefore unable to move either downwards or to one side and as a result they spiral back up the column (often helped by an extraction voltage) until they are outside of the lens field and at this point they are attracted into the upper E-T detector.  In this way a pure secondary electron image may be formed.  

 The high energy BSE are unable to contribute to the image in a double detector instrument unless the manufacturer has a lens design that allows their influence to be incorporated in the upper detector signal.  Converted BSE that strike the lens surfaces producing SE may be incorporated in the final image or excluded depending upon the lens configuration and the specimen position.  In the dual detector system images may be obtained from specimens less than 3mm from the final lens, longer working distances (often 5 to 7mm) offer the opportunity for the BSE influence to contribute in this way.  The constructive influence of BSE should not be discounted when attempting to obtain the maximum information from a specimen area, very often whilst SE offer resolution the BSE offer more information.

 The high resolution performance capabilities of a dual detector system should not inhibit the use of the lower detector or, available on some instruments, the ability to add upper and lower detector signals together.  The lower detector will offer all the facilities and specimen manipulation that is available in a single detector instrument.  Far higher contributions from BSE, therefore higher image depth and image contrast, as well as lower magnifications, will be available from this configuration.

 In short, a double detector instrument offers far wider imaging variations and therefore the ability to extract far more information from the specimen; the ideal system.